C, S Determination in Pure Iron Samples
Application Notes
*Subject to technical modification and errors
Suitable analyzers
Used accessories
- Ceramic crucibles (90150)
- Tungsten (90220)
- Suitable calibration material (NIST or other)
Settings
- Comparator level: 10 mV (low C, S Channel only)
- Minimum time: 45 sec
- Maximum time: 3:00 min
- Integration delay: 6 sec
- Post-waiting: 10 sec
- Pre-purging: 15 sec
- Base line deviation: 5 mV
- Base line time: 30 sec
- T1 – Pre-purging: 5 sec
- T2 – Chamber only: 5 sec
- T3 – Chamber and lance: 1 sec
Sample Preparation
Make sure that the surface of the pure iron is free from contaminations;
otherwise clean the sample with acetone p.a. and let dry at atmosphere.
Pre-heat the crucibles at least 1 h at 1000°C ; let the crucibles cool down
in an exsiccator.
Procedure
– Prepare ELTRA analyzer (e.g. exchange anhydrone, sodium hydoxide,
copper oxide when necessary); clean the combustion tube, brush, heating
shield, dust trap
– Run three warm up samples (e.g. steel samples (92400-3050) with a
minimum weight of 500 mg; add 1.5 g tungsten)
– Calibrate the analyzer with suitable calibration material (NIST or other
(1) Weigh in approx. 500 mg of sample in the ceramic crucibles (90150)
(2) Add 1.7 (±0.1 g) of tungsten (90220)
(3) Place the crucible on the pedestal and start analysis
Repeat step (1) – (3) at least three times;
Mark the results and use the calibration function in the software.
-> Now start with the actual analysis.