Oxygen and Nitrogen Determination in Titanium Samples
Suitable analyzers
- ON 900
- ONH 2000
Used accessories
- Inner graphite crucibles (90180)
- Outer graphite crucibles (90185)
- Nickel basket (90250) or AR 2344
- Suitable calibration material (NIST or other)
Settings
- Comparator level: 20 mV
- Minimum time: 15 sec
- Maximum time: 3:00 min
- Post waiting: 30 sec
- Base line deviation: 200 mV;
- Step: 16 mV; Time: 10 sec
- Mode: continuous
- Outgas
- Time: 45 sec
- Power: 6 kW
- Time
- Purge: 20 sec
- Stability: 45 sec
- Integration delay
- IR cell: 2 sec
- TC cell: 10 sec
- Analyse (1)
- Time: 75 sec
- Power (1): 6 kW
- Power (2): 6 kW
Sample preparation
Make sure that the surface of the titanium is free from contaminations; otherwise clean the sample with acetone p.a. and let dry at atmosphere.
Procedure
- Prepare ELTRA analyzer (exchange anhydrone, sodium hydroxide, copper oxide when necessary), clean furnace, sample drop mechanism, electrode tip
- Run three blanks with empty crucibles
- Calibrate the analyzer with suitable calibration material (NIST or other)
- (1) Fill one empty inner crucible (90180) in one outercrucible (90185) and place them on the electrode tip, close furnace (F2 Button)
- (2) Weigh calibration material and give it into a nickel basket (90250 or AR2344)
- (3) Place nickel basket with calibration material or sample in the sample drop mechanism and start analysis (F5 Button)
After analysis give the inner crucible into waste and fill in a new one. The outer crucibles can be used approximately 10 times.
Repeat steps (1) – (3) at least three times; Mark the results and use the calibration function in the software.
-> Now start with the actual analysis.